Medline Soft Cloth 2-Tube Neonatal Cuffs with Slip Luer Connector
Manufacturer: MEDLINE
Medline Soft Cloth 2-Tube Neonatal Cuffs with Slip Luer Connector
Manufacturer: MEDLINE
- Soft cloth single-patient-use blood pressure cuffs
- Rounded corners and finished edges for optimal comfort
- Hook-and-loop closures ensure a lasting and snug fit for patients
- Meet the American Heart Association recommendations for accuracy
More about this item
Item | DescriptionItem Description | UOM | ||||||
---|---|---|---|---|---|---|---|---|
MDS9751 | Disposable Double-Tube Blood Pressure Cuff with Luer Connector, Neonatal #1 | 10/BX | ||||||
MDS9751H | Disposable Double-Tube Blood Pressure Cuff with Luer Connector, Neonatal #1 | 1 EA | ||||||
MDS9752 | Disposable Double-Tube Blood Pressure Cuff with Luer Connector, Neonatal #2 | 10/BX | ||||||
MDS9752HDiscontinued with Stock | Disposable Double-Tube Blood Pressure Cuff with Luer Connector, Neonatal #2 | 1 EA | ||||||
MDS9753 | Disposable Double-Tube Blood Pressure Cuff with Luer Connector, Neonatal #3 | 10/BX | ||||||
MDS9753H | Disposable Double-Tube Blood Pressure Cuff with Luer Connector, Neonatal #3 | 1 EA | ||||||
MDS9754 | Disposable Double-Tube Blood Pressure Cuff with Luer Connector, Neonatal #4 | 10/BX | ||||||
MDS9755 | Disposable Double-Tube Blood Pressure Cuff with Luer Connector, Neonatal #5 | 10/BX | ||||||
Similar items
More about this item
Find every diagnostic instrument needed to care for patients and residents.
Do you know the real costs of your BP Cuffs?
Research shows that reusable BP cuffs frequently carry live bacteria. The cost both financially and to your patient’s health can be high.
The good news: switch to our disposable cuffs and help reduce cross contamination.
The disposable difference: safety and simplicity.
Our disposable blood pressure cuffs are beneficial for your patients and more efficient for your caregivers. Here’s how:
Choose the right cuff every time.
Find the matching cuff and connector for your BP monitor.
Troubleshooting FAQs for Medline Blood Pressure Cuffs
These cuffs are giving inaccurate readings.
- Medline cuffs are tested to all of the same standards as other leading manufacturers including 1) air leakage testing, 2) dimensional testing, 3) 1,000 inflations cycle testing, and 4) maximum pressure testing.
- The most common issue we see where clinicians are concerned about accuracy is due to sizing. It is important to note that the index line must fall within the range finder box on the BP cuff to ensure that the cuff is the right size for that patient.
- If issues persist, change out the adapter on the cord. Then try testing the cuff on a different machine to see if it is an issue with the machine, the patient, or the cuff. It is important to isolate and eliminate each variable to determine the issue.
These adapters are breaking.
- It is to be expected that there will be a percentage of adapters that break post-conversion, as users get used to the new connection style, especially when going from a push button connector to the HP/bayonet style.
- Do not twist the sheath on the adapter. The sheath should be pulled back when disconnecting the cuff.
- Do not hold the sheath on the adapter when inserting the cuff. Hold the end of the adapter or the tubing when inserting the cuff, so as to not lock the sheath in the upward position.
- We leave extra adapters with BioMed, and they can assist with swapping out the adapters if needed once the Medline team is no longer on site.